Maintenance
60
Lexa | Instructions for Use | MN-111 Med ENG Rev14 | 03/04/2023 | © W&H Sterilization Srl
Monthly or 50-cycle maintenance
CLEANING THE DOOR GASKET AND THE CHAMBER FACE SIDE
Proceed as follows:
1
Clean the door gasket and the outer edge of the chamber with a non-abrasive
cloth moistened with clean water.
Notice
: do not use abrasive products, cutting tools or sharp objects.
1
Rinse with clean water.
Note
: when the seal is new it might be necessary to hold the door gently closed at the
sterilization start.