Diagnostics
70
Lexa | Instructions for Use | MN-111 Med ENG Rev14 | 03/04/2023 | © W&H Sterilization Srl
Code
Description
Actions
4xx
Clean water error (bad quality, clean tank low level, high consumption of water).
Repeat the cycle.
If the problem persists, contact technical service.
Drain and/or refill the clean water tank.
Check the door gasket. Clean or replace it if necessary.
5xx
Check if there are hurdles on the door locking area (chamber rack, loads, objects, …).
Repeat the cycle.
If the problem persists, contact technical service.
Check the door gasket (wrong placed).
Check if the door can move freely without touching the trays or the load when closing.
Switch the sterilizer OFF and ON.
990
The cycle has been aborted by the user.
Re-process the load.