MESSAGES AND ALERTS
Notice
: for any error not listed in this table, contact technical service.
Message/Alert
Description
Action
Fill clean water tank.
There is not enough water in the tank to perform a
cycle.
Fill the water tank as requested.
Drain used water tank.
The used water tank is full.
Drain the water tank as requested.
Please close the door.
The door must be locked, but you didn't close it.
Close the door so it can be locked.
Non-conform water
The clean water quality is bad (conductivity
between 15 and 50 μS/cm).
You may run a cycle but the water must be replaced soon, otherwise the unit will
automatically lock-out to prevent damage.
Poor water quality detected.
Replace with higher quality
water, or equipment damage
may occur.
The clean water quality is very bad (conductivity
more than 50 μS/cm).
Running a cycle is inhibited to prevent damage. Replace the clean water.
Door Gasket must be replaced in
... cycles.
Do you want to replace it now?
These are pre-alerts advising that one of the
consumables has to be replaced within a small
number of cycles.
Tap
if you have the consumable available for replacement.
Tap
if you do not have the consumable in stock and must order one. In this case,
the pre-alert will appear again after some cycles.
See "Maintenance" on page 58.
HEPA Filter must be replaced in
... cycles.
Do you want to replace it now?