Diagnostics
72
Lexa | Instructions for Use | MN-111 Med ENG Rev14 | 03/04/2023 | © W&H Sterilization Srl
Message/Alert
Description
Action
Door Gasket replacement is due.
Do you want to replace it now?
These messages advise that one consumable must
be replaced.
Replace the consumable and tap
to reset the counter (See "Maintenance" on
page 58).
If you do not replace the consumable, press
.
In this case, you may still use the sterilizer but the message will appear again after
some cycles.
CAUTION!
Operating the sterilizer with expired consumables could be
dangerous and could damage the sterilizer.
Have you replaced the door
gasket?
Press YES to reset the counter.
Have you replaced the HEPA
filter?
Press YES to reset the counter.
HEPA Filter replacement is due.
Do you want to replace it now?
Possible leak detected.
Please run Vacuum Test.
Air was detected in the chamber: a vacuum leak is
suspected. The cycle was completed but a vacuum
test is required.
Run a vacuum test.
Call for service if an anomaly is detected.
Please check:
- Not to overload the sterilizer
- The door gasket
If the problem persists contact
the service.
This message advises you that the pressure inside
the chamber did not drop as expected in the first 30
seconds of drying phase.
Check the door gasket correct positioning and make sure do not overload the sterilizer
chamber.
If the problem persists, contact technical service.