Maintenance
62
Lexa | Instructions for Use | MN-111 Med ENG Rev14 | 03/04/2023 | © W&H Sterilization Srl
CLEANING THE CHAMBER FILTERS
Proceed as follows:
1
Allow the sterilization chamber to cool down.
2
Empty the sterilizer chamber by removing the trays and the rack.
3
Remove the filters at the bottom of the chamber (bottom/center) by unscrewing
them.
4
Rinse the filters with tap water.
5
Insert the filters in their original position, screwing them into place.
Note
: tap
to see the animated cleaning procedure.
CLEANING THE EXTERNAL SURFACES OF THE STERILIZER
Proceed as follows:
1
Clean all external sterilizer covers with a slightly damp cloth moistened with
water. For better cleaning results, clean with W&H MC-1000 cleaning solution.
Note
: for cleaning operations, do not dilute the W&H MC-1000 cleaning solution.
Notice
: never use any other disinfectant, detergent or abrasive product, as they
might result aggressive for the external covers and damage them.