10
Lexa | Instructions for Use | MN-111 Med ENG Rev14 | 03/04/2023 | © W&H Sterilization Srl
Introduction
Key program features, including sterilization time, temperature and
recommended load type are listed in the following table:
Program
Type of Load
and Load
weight
Sterilization
Temperature
Sterilization
Time
Drying Time
(recommended)
Pouches
Medium Load
Instruments
and dental
handpieces,
up to 3 lbs
(1.4 kg).
270 °F
(132 °C)
4 minutes
25 minutes
Pouches
Large Load
Instruments
and dental
handpieces,
up to 6 lbs
(2.7 kg).
270 °F
(132 °C)
4 minutes
30 minutes
Wrapped
Cassettes
Instruments
and dental
handpieces,
up to 14 lbs
(6.3 kg).
270 °F
(132 °C)
4 minutes
40 minutes
Program
Type of Load
and Load
weight
Sterilization
Temperature
Sterilization
Time
Drying Time
(recommended)
Low
Temperature
Textiles, up to
4.4 lbs (2 kg),
or
instruments
and dental
handpieces
requiring low
temperature,
up to 5 lbs
(2.3 kg).
250 °F
(121 °C)
30 minutes
30 minutes
Unwrapped
Instruments
and dental
handpieces,
up to 18 lbs
(8.1 kg).
Unwrapped,
naked.
270 °F
(132 °C)
4 minutes
8 minutes
See "Sterilization programs" on page 81 for the full list of key
program features, including sterilization time, temperature, drying
time and recommended load type.
Summary of Contents for MN-111
Page 1: ...MN 111 Instructions for Use MN 111 Med ENG Rev14...
Page 2: ......