Waikato MK5 Technical Manual Download Page 19

 

Mk5 Milk Meter           Compteur a Lait Mark V 
Technical Manual        Manuel Technique 

 
CLEANING       NETTOYAGE

 

19

 

10 February 1995

 

 

6. 

CLEANING PROCEDURES 

 
Under  normal  circumstances  the Waikato  Mk5 
Milk Meter may be satisfactorily washed with a 
`Clean  in  Place'  system,  although  some  herd 
testing  authorities  require  that  the  flask  be 
removed and washed manually. 
 
 
A `manual cleaning' procedure is included as it 
may  be  required  as  part  of  the  Periodic 
Inspection  Section  of  this  manual,  if  the  Milk 
Meters  have  been  used  with  substandard 
cleaning systems. 
 
CAUTION:  
Do  not  use  abrasive  pads  or  powders  on  the 
Milk Meters as these may scratch the surface. 
 
 
Do not use organic solvents or hydrocarbons to 
clean the plastic parts of the Milk Meters.   
 
 
Do  not  use  brushes,  abrasive  pads  or  any 
implement  on  the  Milk  Meter  base  (including 
divider orifice and inlet) or cover. 
 

 

6.1.  CLEAN IN PLACE PROCEDURE 
 
Milking  machine  cleaning  routines  depend  on 
the detergent manufacturer's recommendations 
on  volume,  temperature,  and  concentration. 
The  following  suggested  routine  should 
therefore  be  used  in  conjunction  with  the 
detergent manufacturer's recommendations. 
 
 
Turn  the  flask  tap  upwards  for  all  cleaning 
routines (refer Fig.5.1c) 
 
 
 
6.1.1.    BEFORE MILKING 
 
Flush the milking machine (with Milk Meters) in 
the  normal  manner  using  an  acid  sanitiser 
solution 

at 

recommended 

concentration, 

temperature and volume. 
 
This step must be carried out prior to milking if  
THE  MILK  METER  HAS  NOT  BEEN  the  Milk 
Meter has not been used for several days. 
 
 
 

 
6. METHODES DE NETTOYAGE 
 
Dans  la  pratique,  le  compteur  à  lait  Waikato 
Mk5 peut être nettoyé de manière satisfaisante 
avec  un  dispositif  de  nettoyage  en  place; 
néanmoins,  certains  organismes  de  contrôle 
exigent  que  la  réserve  soit  retirée  et  lavée 
manuellement. 
 
Une méthode de nettoyage manuel est décrite 
ci-après  car  pouvant  rentrer  dans  le  contrôle 
périodique  des  compteurs  défini  dans  ce 
manuel,  dans  le  cas  de  dispositifs  de 
nettoyage en place non standard. 
 
ATTENTION : 
Ne  pas  utiliser  de  poudres  ni  chiffons  abrasifs 
pour  le  nettoyage  des  compteurs  car  ceux-ci 
peuvent en rayer la surface. 
 
Ne  pas  utiliser  de  solvants  organiques  ni 
d’hydrocarbures  pour  nettoyer  les  parties 
plastiques des compteurs à lait. 
 
Ne  pas  utiliser  de  brosses  ou  de  chiffons 
abrasifs  pour  nettoyer  la  base  du  compteur  ni 
même  l’orifice  de  prélèvement,  l’entrée  ou  le 
couvercle. 
 
 
6.1. METHODE DE NETTOYAGE EN PLACE 
 
Les  méthodes  de  nettoyage  des  machines  à 
traire  dépendent  des  recommandations  du 
fabricant  pour  la  quantité  de  détergent  à 
employer, sa température optimum d’emploi et 
aussi  sa  concentration.  La  méthode  de 
nettoyage  décrite  ci-après  devra  donc  être 
appliquée  en  fonction  des  recommandations 
du fabricant. 
 
Tourner le robinet de la réserve avec sa sortie 
orientée vers le haut pour toutes les méthodes 
de nettoyage (voir Fig 5.1c). 
 
 
6.1.1. AVANT LA TRAITE 
 
Faire  circuler  une  solution  acide  en  volume, 
concentration  et  à  température  recommandés 
pour  le  pré-nettoyage  de  la  machine  à  traire 
avec compteurs à lait en place. 
 
Cette  première  étape  doit  être  effectuée  avant 
toute  traite  quand  les  compteurs  à  lait  n’ont 
pas été utilisés pendant plusieurs jours. 
 

Summary of Contents for MK5

Page 1: ...MARK 5 MILK METER TECHNICAL MANUAL COMPTEUR A LAIT MARK V MANUEL TECHNIQUE C 1993 1995 InterAg Hamilton New Zealand...

Page 2: ...OCEDURE 18 6 1 1 BEFORE MILKING 18 6 1 2 IMMEDIATELY AFTER USE 19 6 1 3 REVERSE FLOW CLEANING SYSTEMS 19 6 1 4 ALKALINE DETERGENTS 19 6 2 MANUAL CLEANING PROCEDURE 20 7 ROUTINE MAINTENANCE 21 8 MILKIN...

Page 3: ...NT LA TRAITE 18 6 1 2 IMMEDIATEMENT APRES UTILISATION 19 6 1 3 SYSTEMES DE NETTOYAGE PAR FLUX INVERSE 19 6 1 4 DETERGENTS ALCALINS 19 6 2 METHODE DE NETTOYAGE MANUEL 20 7 MAINTENANCE DE ROUTINE 21 8 C...

Page 4: ...ents 1 INTRODUCTION Le compteur lait Waikato est un dispositif de mesure proportionnel de la quantit de lait produite situ entre le tuyau lait de chaque poste et la ligne de lait Le compteur lait s pa...

Page 5: ...cision selon les crit res de ICAR International Commitee for Animal Recording et du DHIA Homologu ind pendamment pour utilisation dans quinze des plus grands pays laitiers du monde Possibilit s de me...

Page 6: ...ait D par lequel monte le lait venant de la griffe avec chaque pulsation de la machine traire Des ailettes en spirale C situ es dans la partie haute du tube d admission impriment une rotation au lait...

Page 7: ...place 3 2 CAPACITES DE LA RESERVE GRADUEE a R serve pour 30kg Homologu e ICAR moul e en une seule pi ce b R serve pour 42kg Couvercle de r serve et joint amovibles pour nettoyage lorsque n cessaire L...

Page 8: ...Fig 3 2c for flat surfaces or piping Wedge Mounting Set Fig 3 2 d and e for flat surfaces or piping Hanger bracket Fig 3 2 f off piping 3 4 POSSIBILITES DE MONTAGE Le compteur lait peut tre mont de m...

Page 9: ...che suspendue Wedge mounting kit V Packer U Bolt kit Wedge mounting bracket Spine Wedge block Stem Wedge block kit Hanger bracket 38050141 38050265 38050186 38050263 38050182 37112000 38050706 Fig 3 2...

Page 10: ...ermany Holland Sweden Ireland New Zealand Korea Scotland Wales South Africa Denmark United States of America 3 5 DIFFERENCES DE MARQUES Le compteur lait Waikato Mk5 est distribu dans la plupart des pa...

Page 11: ...5 38050505 38050505 R Stainless Steel Connector 38050704 38050704 38050704 IR S S Connector Assembly 38050166 38050166 38050166 K Hose 38050506 38050506 38050506 L 2 Flask 30kg 38050164 L 3 Flask 42kg...

Page 12: ...0505 38050505 38050505 R Raccord inox 38050704 38050704 38050704 IR Raccord inox complet 38050166 38050166 38050166 K Tuyau 38050506 38050506 38050506 L 2 R serve 30kg 38050164 L 3 R serve 42kg 380501...

Page 13: ...ovac and the Duovac switched to manual 4 1 EMPLACEMENT voir Fig 4 1 Le compteur lait est install partir du tuyau lait entre le faisceau de traite et la tuyauterie de la ligne de lait Il doit tre insta...

Page 14: ...chnique INSTALLATION 14 10 February 1995 High Line Mid Line Curb Low Line Cow Floor ACR Vacuum Tap Fig 4 1 Installation of Milk Meters Ligne Haute Ligne Centrale Decrochage Automatique Ligne Basse Bor...

Page 15: ...odified ball cage assembly fitted 4 5 CONVERSION DU COMPTEUR WAIKATO DE 30kg 65lb 42kg 92lb A l exception de la r serve des bouchons de r serve type ouvert ou ferm et de la chambre de la bille inox to...

Page 16: ...e la r serve avant utilisation 2 Fermer le robinet de la r serve en le tournant jusque sa position horizontale voir Fig 5 1a Bien enfoncer le robinet dans son logement afin d assurer sa position 3 Pla...

Page 17: ...r releases sample via tap outlet POSITION PRISE D ECHANTILLON Sortie du robinet orient e vers le bas pour l admission d air dans la r serve pour agiter le lait En appuyant sur l onglet de mise hors pr...

Page 18: ...as per the cleaning procedure section of this manual 5 2 UTILISATION DU COMPTEUR AVEC RESERVE A FOND FERME Pour une lecture de la r serve gradu e et une prise d chantillon pour analyse ult rieur hors...

Page 19: ...eter has not been used for several days 6 METHODES DE NETTOYAGE Dans la pratique le compteur lait Waikato Mk5 peut tre nettoy de mani re satisfaisante avec un dispositif de nettoyage en place n anmoin...

Page 20: ...fect of this 6 1 2 IMMEDIATEMENT APRES UTILISATION 1 Rincer l installation de traite avec les compteurs lait en place en utilisant de l eau soit froide soit ti de jusqu ce que l eau vacu e sorte incol...

Page 21: ...shake to clean A sponge may be used to remove stubborn deposits 3 Rinse with ample fresh warm water 6 2 METHODE DE NETTOYAGE MANUEL A utiliser comme partie int grante de la M thode de R vision P riodi...

Page 22: ...ction and Periodic Test at least once annually 7 MAINTENANCE DE ROUTINE De par sa conception le compteur lait ne requiert qu un minimum de maintenance Toutefois et comme faisant partie de ses connaiss...

Page 23: ...eters are fitted Note On no account should the vacuum level ever be set greater than 50kPa 15 Hg 8 CONDITIONS REQUISES PAR LA MACHINE A TRAIRE Pour assurer un bon tat physique et sanitaire des vaches...

Page 24: ...exit hole in the cup Any abnormal restrictions such as these can affect Milk Meter accuracy 8 3 ENTREE D AIR AU NIVEAU DE LA GRIFFE L admission d air au niveau de la griffe est essentiel afin d assur...

Page 25: ...ycerine on the mating surfaces would help here 9 PROBLEMES DE FONCTIONNEMENT 9 1 INCIDENTS 9 1 1 BULLES D AIR DANS LE LAIT DE LA RESERVE Cause robinet trois voies us sec ou mal plac Rem de remplacer h...

Page 26: ...tterfat samples not taken in duplicate 9 2 PROBLEMES DE PRECISION 9 2 1 VERIFICATION DE LA PRECISION DES COMPTEURS PAR DEUX METHODES a Comparer la production individuelle d une vache en pesant s par m...

Page 27: ...ease cleaning solution volume Flow rate Concentration Temperature A cleaning system that is clearly inadequate would require redesigning for efficient cleaning 9 3 PROBLEMES DE NETTOYAGE Si le syst me...

Page 28: ...solvents to clean Milk Meter parts 10 REVISION PERIODIQUE Au moins une fois par an tous les compteurs lait doivent tre contr l s selon la m thode suivante voir Fig 3 3 pour le rep rage des pi ces 10...

Page 29: ...and Recalibration Contr ler soigneusement les pi ces suivantes du compteur selon leur degr d usure tat des surfaces coupures fentes et rayures Robinet trois voies Tuyaux d entr e et de sortie Bouchon...

Page 30: ...source hose B 5 Volumetric flasks or accurate scales will be required 11 METHODE DE TEST PERIODIQUE Conjointement avec la r vision p riodique tous les compteurs doivent tre test s pour leur pr cision...

Page 31: ...er and repeat the procedure to obtain two results per Meter 11 2 METHODE POUR TEST DE BASE 1 Remplir le seau D avec 16 litres d eau 16kg ou 35 24lb qui sera le fluide de travail 2 Ouvrir le robinet d...

Page 32: ...se Tap C Supply Pail E fitted to lowest point Flow Restrictor R duction de d bit Flow Control Tube Tube de contr le F Air Admission 600mm from Milk Meter inlet Fig 11 1 Basic Milk Meter Test Rig To Va...

Page 33: ...tted to a certified service agent for repair and recalibration Note The periodic test procedure should be regularly followed to give confidence in Milk Meter results 11 3 ANALYSE DES RESULTATS Calcule...

Page 34: ...for any points of non compliance with these recommended procedures 12 REPARATIONS ET METHODE D ETALONNAGE Ce paragraphe traite de la r paration et de l talonnage du compteur lait Waikato qui ne peut t...

Reviews: