Waikato MK5 Technical Manual Download Page 10

 

Mk5 Milk Meter           Compteur a Lait Mark V 
Technical Manual        Manuel Technique 

 
          DESCRIPTION 

 

10

 

10 February 1995

 

 

3.5.  LABELLING VARIATIONS 
 
The  Waikato  MK5  Milk  Meter  is  marketed  in 
most of the dairying countries of the world and 
may  be  labelled  differently  depending  on  the 
distributor in each country. 
 
In each case the basic Milk Meter is the same.  
The  only  structural  differences  between  labels 
are the presence or absence of fittings such as 
spine assembly,  mounting set, or packaging. 

 

3.6.  QUALITY ASSURANCE 
 
Each  Milk  Meter  is  manufactured  and 
assembled  to  strict  quality  control  criteria  and 
each  Milk  Meter  is  tested  under  simulated 
milking conditions.   
 
 
This  ensures  that  Milk  Meters  on  leaving  the 
factory  are  accurate  to  within  plus  or  minus 
1.3% at a yield of 15kg.   
 
 
It  is  recommended  that  each  Milk  Meter  be 
subjected  to  at  least  an  annual  inspection  and 
test  procedure  as  detailed  in  the  Periodic 
Inspection section of this manual. 
 
If  repair  or  recalibration  is  required  this  should 
be  carried  out  only  by  suitably  trained 
personnel. For example, in the United States of 
America,  only  at  DHIA  Certified  Repair  and 
Recalibration Centres. 
 

3.7.  WAIKATO 

 

MK5 

MILK 

METER  

APPROVAL 
 
The  Waikato  Mk5  Milk  Meter  has  been 
provisionally  approved  by  ICAR  (International 
Committee  for  Animal  Recording).  Also  the 
Milk  Meter  has  the  approval  of  individual 
testing authorities in the following countries: 
 
 
Australia 

 

Canada   

 

Japan 

 

England 

France 

 

Italy 

 

Germany 

 

Holland   

Sweden 

 

Ireland   

New Zealand   

Korea 

Scotland 

 

Wales   

South Africa   

Denmark 

 

United States of America 

 

 

3.5  DIFFERENCES DE MARQUES 
 
Le  compteur  à  lait  Waikato  Mk5  est  distribué 
dans la plupart des pays producteurs de lait et 
peur  être  l’objet  de  différentes  marques  selon 
les importateurs. 
 
Dans  tous  les  cas  le  compteur  à  lait  de  base 
reste le même. Les seules différences notables 
entre  les  marques  résident  dans  la  présence 
ou  l’absence  de  supports,  kits  de  montage  ou 
emballage. 
 
3.6  GARANTIE DE QUALITE 
 
Chaque  compteur  à  lait  est  fabriqué  et 
assemblé  selon  des  critères  de  qualité  très 
strictes  et  chaque  compteur  à  lait  est  testé 
individuellement  en  simulant  les  conditions  de 
la traite. 
 
Cela permet d’estimer que les compteurs à lait 
sortant  d’usine  possèdent  une  précision  de 
plus  ou  moins  1,3%  pour  un  rendement  de 
15kg. 
 
Il  est  fortement  recommandé  d’effectuer  une 
révision annuelle de chaque compteur à lait et 
d’effectuer les tests de contrôle comme définis 
dans ce manuel technique. 
 
Si  une  réparation  ou  un  étalonnage  sont 
requis,  ceux  ci  devront  être  effectués  par  un 
personnel qualifié. A titre d’exemple, aux USA, 
seuls 

les 

personnels 

des 

Centres 

de 

Réparations  et  d’étalonnage  DHIA  sont 
autorisés pour ce travail. 
 
3.7  HOMOLOGATION  DU  COMPTEUR  A 
LAIT WAIKATO MK5 
 
Le  compteur  à  lait  Waikato  Mk5  a  été 
provisoirement 

approuvé 

par 

ICAR 

(International Commitee for Animal Recording). 
Le  compteur  à  lait  possède  également 
l’homologation 

de 

centres 

de 

contrôle 

individuels dans les pays suivants : 
 
Australie 

 

Canada   

Japon 

 

Angleterre 

France 

 

Italie 

Allemagne 

 

Hollande 

Suède 

 

Irlande   

Nouvelle Zélande 

Corée 

Ecosse 

 

Pays de Galles 

Afrique du Sud 

Danemark 

USA  

 

 

Summary of Contents for MK5

Page 1: ...MARK 5 MILK METER TECHNICAL MANUAL COMPTEUR A LAIT MARK V MANUEL TECHNIQUE C 1993 1995 InterAg Hamilton New Zealand...

Page 2: ...OCEDURE 18 6 1 1 BEFORE MILKING 18 6 1 2 IMMEDIATELY AFTER USE 19 6 1 3 REVERSE FLOW CLEANING SYSTEMS 19 6 1 4 ALKALINE DETERGENTS 19 6 2 MANUAL CLEANING PROCEDURE 20 7 ROUTINE MAINTENANCE 21 8 MILKIN...

Page 3: ...NT LA TRAITE 18 6 1 2 IMMEDIATEMENT APRES UTILISATION 19 6 1 3 SYSTEMES DE NETTOYAGE PAR FLUX INVERSE 19 6 1 4 DETERGENTS ALCALINS 19 6 2 METHODE DE NETTOYAGE MANUEL 20 7 MAINTENANCE DE ROUTINE 21 8 C...

Page 4: ...ents 1 INTRODUCTION Le compteur lait Waikato est un dispositif de mesure proportionnel de la quantit de lait produite situ entre le tuyau lait de chaque poste et la ligne de lait Le compteur lait s pa...

Page 5: ...cision selon les crit res de ICAR International Commitee for Animal Recording et du DHIA Homologu ind pendamment pour utilisation dans quinze des plus grands pays laitiers du monde Possibilit s de me...

Page 6: ...ait D par lequel monte le lait venant de la griffe avec chaque pulsation de la machine traire Des ailettes en spirale C situ es dans la partie haute du tube d admission impriment une rotation au lait...

Page 7: ...place 3 2 CAPACITES DE LA RESERVE GRADUEE a R serve pour 30kg Homologu e ICAR moul e en une seule pi ce b R serve pour 42kg Couvercle de r serve et joint amovibles pour nettoyage lorsque n cessaire L...

Page 8: ...Fig 3 2c for flat surfaces or piping Wedge Mounting Set Fig 3 2 d and e for flat surfaces or piping Hanger bracket Fig 3 2 f off piping 3 4 POSSIBILITES DE MONTAGE Le compteur lait peut tre mont de m...

Page 9: ...che suspendue Wedge mounting kit V Packer U Bolt kit Wedge mounting bracket Spine Wedge block Stem Wedge block kit Hanger bracket 38050141 38050265 38050186 38050263 38050182 37112000 38050706 Fig 3 2...

Page 10: ...ermany Holland Sweden Ireland New Zealand Korea Scotland Wales South Africa Denmark United States of America 3 5 DIFFERENCES DE MARQUES Le compteur lait Waikato Mk5 est distribu dans la plupart des pa...

Page 11: ...5 38050505 38050505 R Stainless Steel Connector 38050704 38050704 38050704 IR S S Connector Assembly 38050166 38050166 38050166 K Hose 38050506 38050506 38050506 L 2 Flask 30kg 38050164 L 3 Flask 42kg...

Page 12: ...0505 38050505 38050505 R Raccord inox 38050704 38050704 38050704 IR Raccord inox complet 38050166 38050166 38050166 K Tuyau 38050506 38050506 38050506 L 2 R serve 30kg 38050164 L 3 R serve 42kg 380501...

Page 13: ...ovac and the Duovac switched to manual 4 1 EMPLACEMENT voir Fig 4 1 Le compteur lait est install partir du tuyau lait entre le faisceau de traite et la tuyauterie de la ligne de lait Il doit tre insta...

Page 14: ...chnique INSTALLATION 14 10 February 1995 High Line Mid Line Curb Low Line Cow Floor ACR Vacuum Tap Fig 4 1 Installation of Milk Meters Ligne Haute Ligne Centrale Decrochage Automatique Ligne Basse Bor...

Page 15: ...odified ball cage assembly fitted 4 5 CONVERSION DU COMPTEUR WAIKATO DE 30kg 65lb 42kg 92lb A l exception de la r serve des bouchons de r serve type ouvert ou ferm et de la chambre de la bille inox to...

Page 16: ...e la r serve avant utilisation 2 Fermer le robinet de la r serve en le tournant jusque sa position horizontale voir Fig 5 1a Bien enfoncer le robinet dans son logement afin d assurer sa position 3 Pla...

Page 17: ...r releases sample via tap outlet POSITION PRISE D ECHANTILLON Sortie du robinet orient e vers le bas pour l admission d air dans la r serve pour agiter le lait En appuyant sur l onglet de mise hors pr...

Page 18: ...as per the cleaning procedure section of this manual 5 2 UTILISATION DU COMPTEUR AVEC RESERVE A FOND FERME Pour une lecture de la r serve gradu e et une prise d chantillon pour analyse ult rieur hors...

Page 19: ...eter has not been used for several days 6 METHODES DE NETTOYAGE Dans la pratique le compteur lait Waikato Mk5 peut tre nettoy de mani re satisfaisante avec un dispositif de nettoyage en place n anmoin...

Page 20: ...fect of this 6 1 2 IMMEDIATEMENT APRES UTILISATION 1 Rincer l installation de traite avec les compteurs lait en place en utilisant de l eau soit froide soit ti de jusqu ce que l eau vacu e sorte incol...

Page 21: ...shake to clean A sponge may be used to remove stubborn deposits 3 Rinse with ample fresh warm water 6 2 METHODE DE NETTOYAGE MANUEL A utiliser comme partie int grante de la M thode de R vision P riodi...

Page 22: ...ction and Periodic Test at least once annually 7 MAINTENANCE DE ROUTINE De par sa conception le compteur lait ne requiert qu un minimum de maintenance Toutefois et comme faisant partie de ses connaiss...

Page 23: ...eters are fitted Note On no account should the vacuum level ever be set greater than 50kPa 15 Hg 8 CONDITIONS REQUISES PAR LA MACHINE A TRAIRE Pour assurer un bon tat physique et sanitaire des vaches...

Page 24: ...exit hole in the cup Any abnormal restrictions such as these can affect Milk Meter accuracy 8 3 ENTREE D AIR AU NIVEAU DE LA GRIFFE L admission d air au niveau de la griffe est essentiel afin d assur...

Page 25: ...ycerine on the mating surfaces would help here 9 PROBLEMES DE FONCTIONNEMENT 9 1 INCIDENTS 9 1 1 BULLES D AIR DANS LE LAIT DE LA RESERVE Cause robinet trois voies us sec ou mal plac Rem de remplacer h...

Page 26: ...tterfat samples not taken in duplicate 9 2 PROBLEMES DE PRECISION 9 2 1 VERIFICATION DE LA PRECISION DES COMPTEURS PAR DEUX METHODES a Comparer la production individuelle d une vache en pesant s par m...

Page 27: ...ease cleaning solution volume Flow rate Concentration Temperature A cleaning system that is clearly inadequate would require redesigning for efficient cleaning 9 3 PROBLEMES DE NETTOYAGE Si le syst me...

Page 28: ...solvents to clean Milk Meter parts 10 REVISION PERIODIQUE Au moins une fois par an tous les compteurs lait doivent tre contr l s selon la m thode suivante voir Fig 3 3 pour le rep rage des pi ces 10...

Page 29: ...and Recalibration Contr ler soigneusement les pi ces suivantes du compteur selon leur degr d usure tat des surfaces coupures fentes et rayures Robinet trois voies Tuyaux d entr e et de sortie Bouchon...

Page 30: ...source hose B 5 Volumetric flasks or accurate scales will be required 11 METHODE DE TEST PERIODIQUE Conjointement avec la r vision p riodique tous les compteurs doivent tre test s pour leur pr cision...

Page 31: ...er and repeat the procedure to obtain two results per Meter 11 2 METHODE POUR TEST DE BASE 1 Remplir le seau D avec 16 litres d eau 16kg ou 35 24lb qui sera le fluide de travail 2 Ouvrir le robinet d...

Page 32: ...se Tap C Supply Pail E fitted to lowest point Flow Restrictor R duction de d bit Flow Control Tube Tube de contr le F Air Admission 600mm from Milk Meter inlet Fig 11 1 Basic Milk Meter Test Rig To Va...

Page 33: ...tted to a certified service agent for repair and recalibration Note The periodic test procedure should be regularly followed to give confidence in Milk Meter results 11 3 ANALYSE DES RESULTATS Calcule...

Page 34: ...for any points of non compliance with these recommended procedures 12 REPARATIONS ET METHODE D ETALONNAGE Ce paragraphe traite de la r paration et de l talonnage du compteur lait Waikato qui ne peut t...

Reviews: