Trouble shooting
NOTE!
Contact the service department if you cannot rectify this error.
Error
Cause
Remedy
The system does not start
No supply of power
Power pack or system control
defect
Provide power
Replace the power pack or
system control
Dispensing not possible/ storage
tank is empty
Feedwater supply is closed
Feedwater and concentrate
water connections are mixed
up
Feedwater pressure < 1 bar
Sterile filter blocked
RO/pretreatment cartridge
blocked.
Open the feedwater supply line
Replace connections
Increase feedwater pressure
Replace the sterile filter (see
“Replacing the 0.2 µm final
filter” on page 61)
Install new RO/pretreatment
cartridge (see “Replacing the
RO/pretreatment” on page 52
or check RO/pretreatment
cartridge connections.
Resistance < 18.2 M
Ω
·cm/ 0.055
µS/cm
Ultrapure cartridge is
exhausted
Air in the ultrafilter because of
empty tank
Check feedwater quality
Insert new ultrapure cartridge
Wait until the tank is again
filled and a rinse is triggered
Control panel non responsive
Microprocessor locked up
PCB error
Unplug the mains plug for 5
seconds.
Contact the VWR service
department
Summary of Contents for Puranity TU 12 UV+
Page 6: ...vi Puranity TU VWR...
Page 18: ...Extend of delivery 10 Puranity TU VWR Chapter 3...
Page 20: ...Intended use of the device 12 Puranity TU VWR Chapter 4...
Page 44: ...Flow chart 36 Puranity TU VWR Chapter 8...
Page 46: ...How the Puranity TU system functions 38 Puranity TU VWR Chapter 9...
Page 72: ...Equipment disposal 64 Puranity TU VWR Chapter 14...
Page 76: ...Consumables 68 Puranity TU VWR Chapter 16...
Page 77: ...VWR Puranity TU 69 Terminal assignment...
Page 78: ...Terminal assignment 70 Puranity TU VWR Chapter 17...
Page 80: ...Maintenance record 72 Puranity TU VWR Chapter 18...
Page 84: ...76 Puranity TU VWR Index...