VWR
Puranity TU+ |
45
System control |
Chapter 12
Setting the limiting value for conductivity
NOTE!
The limiting value can only be set in µS/cm and is factory adjusted to a value of 0.1
µS/cm.
Setting range: 0.055 - 30 µS/cm, basic setting: 0.1 µS/cm
Step
Action
Figure
1
Press the “Menu”
button twice to enable
the limiting value for
the conductivity to be
set (red LED lights up).
2
Use the “Up/quit” and
“down” button to
increase or reduce the
value. With the “Enter”
button you switch to
the next digit. After
finishing adjustment
press “Enter” to
confirm the value.
3
Press the “Menu“
button twice to return
to the operating mode.
Display shows
Cond. Temp.
Status Sensor
Second line
Display
0.10
Limit value
or
Display shows
0.10
0.10
Display shows