| Maintenance
50
|
Puranity TU+
VWR
Chapter 13
Please note that the lifetime of the wear parts is a direct function of the quality of the feedwater
and the daily volume of water that is used.
Replacing the ultrapure cartridge
NOTE!
Replace the ultrapure cartridge with a new one when the ultrapure water conductivity is purity
dropped below acceptable levels or the replacement interval is due.
Material
Flow chart no.
Cat. No.:
Interval
Other problems
Ultrapure
cartridge
F1
171-1141
12 months
Or when the pure water limit value is
exceeded, whichever occurs first. Bacteria
growth may occur in the resin when the
system has been in use over an extended
period.
Sterile filter 0.2
µm
F2
171-1105
12 months
Or the flow rate is markedly slower
Ultrafiltration
membrane (UF)
F3
171-1106
24 months
Or if there is endotoxin breakthrough in
product water or when the water flow rate is
markedly slowler.
UV lamp
UV1
171-1108
24 months
Or when UV lamp does not light.
RO/pretreatment
cartridge
F6
171-1142
171-1143
171-1165
12 months
Or when the ultrapure cartridge has shorter
life time than excepted.
Step
Action
Figure
1
Switch the system off (unplug main
power supply).
2
Remove the cover from the ultrapure
cartridge by press the snapper and
pull the cover toward the front to
remove it.
Snapper
Cover
Summary of Contents for Puranity TU 12 UV+
Page 6: ...vi Puranity TU VWR...
Page 18: ...Extend of delivery 10 Puranity TU VWR Chapter 3...
Page 20: ...Intended use of the device 12 Puranity TU VWR Chapter 4...
Page 44: ...Flow chart 36 Puranity TU VWR Chapter 8...
Page 46: ...How the Puranity TU system functions 38 Puranity TU VWR Chapter 9...
Page 72: ...Equipment disposal 64 Puranity TU VWR Chapter 14...
Page 76: ...Consumables 68 Puranity TU VWR Chapter 16...
Page 77: ...VWR Puranity TU 69 Terminal assignment...
Page 78: ...Terminal assignment 70 Puranity TU VWR Chapter 17...
Page 80: ...Maintenance record 72 Puranity TU VWR Chapter 18...
Page 84: ...76 Puranity TU VWR Index...