VWR
Puranity TU+ |
51
Maintenance |
Chapter 13
3
Press on the two quick connectors at
the inlet and outlet of the ultrapure
cartridge and remove the used
cartridge from the system.
4
We recommend performing a cleaning
when an existing ultrapure cartridge is
replaced.
NOTE!
Refer to section “Cleaning” on page 53 for an explanation of
how to perform cleaning.
5
Locate the new ultrapure cartridge and
insert the cartridge into the system.
6
Insert the quick connectors into the
connecting points on the ultrapure
cartridge. When you hear an audible
click you can be sure that the quick
connectors have been inserted
correctly.
NOTE!
The quick-connectors are attached
to the system in such a manner so it
is to prevent any confusion
(switching).
7
Put the cover for the ultrapure
cartridge back in place and switch the
system on (connect the main power
supply).
8
NOTE!
Discharge the first 5 liters.
Step
Action
Figure
Ultrapure
cartridge
Quick
connectors
Ultrapure
cartridge
Summary of Contents for Puranity TU 12 UV+
Page 6: ...vi Puranity TU VWR...
Page 18: ...Extend of delivery 10 Puranity TU VWR Chapter 3...
Page 20: ...Intended use of the device 12 Puranity TU VWR Chapter 4...
Page 44: ...Flow chart 36 Puranity TU VWR Chapter 8...
Page 46: ...How the Puranity TU system functions 38 Puranity TU VWR Chapter 9...
Page 72: ...Equipment disposal 64 Puranity TU VWR Chapter 14...
Page 76: ...Consumables 68 Puranity TU VWR Chapter 16...
Page 77: ...VWR Puranity TU 69 Terminal assignment...
Page 78: ...Terminal assignment 70 Puranity TU VWR Chapter 17...
Page 80: ...Maintenance record 72 Puranity TU VWR Chapter 18...
Page 84: ...76 Puranity TU VWR Index...