| Maintenance
56
|
Puranity TU+
VWR
Chapter 13
Changing the ultrafilter
NOTE!
You will need the following tools for replacing the ultra-filter:
Open-end wrench, size 17, Phillips screwdriver and Teflon tape.
Step
Action
Figure
1
Switch the system off (unplug main power
supply).
2
a.
Remove the cover from the system.
b.
Take out the ultrapure cartridge and
RO/pretreatment cartridge (see
“Replacing the ultrapure cartridge”
on page 50 and “Replacing the
RO/pretreatment” on page 52).
c.
Use the Phillips screwdriver to
unscrew the screws from the
bracket for the UV lamp and pull the
UV assembly out toward the front.
d.
Unscrew the threaded connection
screws by hand on the d1/4“ fittings
of the ultra-filter and pull the tubes
out. Disconnect the fitting on the
upper side from the ultra-filter. Twist
the ultrafliter to turn out the
measuring cell from the upper side
from the ultrafilter.
NOTE!
Ensure that the white O-rings on the 1/4“
inch tubes are not lost in the process. You
will need these again for re-attaching the
tube.
c.)
d.)
UV lamp
Mounting plate
Threaded
connection
fitting
Fitting
Screws
Ultrafilter
Measuring cell
Summary of Contents for Puranity TU 12 UV+
Page 6: ...vi Puranity TU VWR...
Page 18: ...Extend of delivery 10 Puranity TU VWR Chapter 3...
Page 20: ...Intended use of the device 12 Puranity TU VWR Chapter 4...
Page 44: ...Flow chart 36 Puranity TU VWR Chapter 8...
Page 46: ...How the Puranity TU system functions 38 Puranity TU VWR Chapter 9...
Page 72: ...Equipment disposal 64 Puranity TU VWR Chapter 14...
Page 76: ...Consumables 68 Puranity TU VWR Chapter 16...
Page 77: ...VWR Puranity TU 69 Terminal assignment...
Page 78: ...Terminal assignment 70 Puranity TU VWR Chapter 17...
Page 80: ...Maintenance record 72 Puranity TU VWR Chapter 18...
Page 84: ...76 Puranity TU VWR Index...