| Maintenance
62
|
Puranity TU+
VWR
Chapter 13
Autoclaving the 0.2 µm final filter
NOTE!
The lifetime of the 0.2 µm final filter can be prolonged by sterilizing it. Proceed as follows to
autoclave the sterile filter.
Step
Action
Figure
1
Rotate the used 0.2 µm final filter in a
clockwise direction to remove it.
2
Use an autoclave to sterilize the 0.2 µm final
filter in a steam autoclave at 121C° for 30
minutes. You can repeat the sterilization
process for the sterile filter up to 5 times.
When sterilization is completed, screw the 0.2
µm final filter back into the 1/4“ thread on the
sampling tap (see “Replacing the 0.2 µm final
filter” on page 61).
NOTE!
If you wish to remove purified water and the flow rate
is too low, this is an indication that the sterile filter is
clogged. In this case, read the description in Chapter
15 “Trouble shooting” on page 77, or replace the 0.2
µm final filter with a new one.
1/4“ thread
0.2 µm final filter
Summary of Contents for Puranity TU 12 UV+
Page 6: ...vi Puranity TU VWR...
Page 18: ...Extend of delivery 10 Puranity TU VWR Chapter 3...
Page 20: ...Intended use of the device 12 Puranity TU VWR Chapter 4...
Page 44: ...Flow chart 36 Puranity TU VWR Chapter 8...
Page 46: ...How the Puranity TU system functions 38 Puranity TU VWR Chapter 9...
Page 72: ...Equipment disposal 64 Puranity TU VWR Chapter 14...
Page 76: ...Consumables 68 Puranity TU VWR Chapter 16...
Page 77: ...VWR Puranity TU 69 Terminal assignment...
Page 78: ...Terminal assignment 70 Puranity TU VWR Chapter 17...
Page 80: ...Maintenance record 72 Puranity TU VWR Chapter 18...
Page 84: ...76 Puranity TU VWR Index...