NI-9231
Page 1: ...NI 9231...
Page 2: ...Update 6 Reverification 6 Accuracy under Calibration Conditions 6 Worldwide Support and Services 7 Software Calibrating the NI 9231 requires the installation of NI DAQmx 18 1 or later on the calibrat...
Page 3: ...AQmx NET Help Support for Visual Studio Reference information for NI DAQmx NET methods and NI DAQmx NET properties key concepts and a C enum to NET enum mapping table Test Equipment The following tabl...
Page 4: ...ement circuitry is at a stable operating temperature Initial Setup Complete the following steps to set up the NI 9231 1 Install NI DAQmx 2 Make sure that the NI cDAQ 9178 power source is not connected...
Page 5: ...s the NI 9231 ground connection 3 Set the calibrator voltage to a Test Point value indicated in the NI 9231 Verification Test Limits table 4 Set the calibrator to Operate mode OPR 5 Acquire and averag...
Page 6: ...stment The following performance adjustment procedure describes the sequence of operation required to adjust the NI 9231 Accuracy Adjustment Complete the following procedure to adjust the accuracy per...
Page 7: ...an external calibration setting the C Series calibration temperature and closing the external calibration Reverification Repeat the Verification section to determine the As Left status of the device...
Page 8: ...t ni com services for information about the services NI offers Visit ni com register to register your NI product Product registration facilitates technical support and ensures that you receive importa...
Page 9: ...ments EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy...