6. Compare the averaged samples to the limits in the 

NI 9231 Verification Test Limits

 table

to verify the input accuracy. If the value is within these limits, the device passes the test.

7. Repeat steps 3 through 7 for each test point in the 

NI 9231 Verification Test Limits

 table.

8. Set the calibrator to Standby mode (STBY).
9. Repeat steps 1 through 8 for each channel on the NI 9231.

Table 4.

 NI 9231 Verification Test Limits

Test Point

2-Year Limits

Location

Value (V)

Lower Limit (VDC)

Upper Limit (VDC)

 

 

 

 

Minimum

-4.9

-4.907121

-4.892879

Mid-scale

0

-0.001565

0.001565

Maximum

4.9

4.892879

4.907121

Note

 

The test limits listed in the 

NI 9231 Verification Test Limits

 table are derived

using values from the 

NI 9231 Accuracy Under Calibration Conditions

 table.

Adjustment

The following performance adjustment procedure describes the sequence of operation required

to adjust the NI 9231.

Accuracy Adjustment

Complete the following procedure to adjust the accuracy performance of the NI 9231.
1. Set the calibrator to Standby mode (STBY).
2. Connect channel 0 of the NI 9231 to the calibrator as shown in the 

Accuracy Verification

section.

3. Adjust the NI 9231.

a) Initialize a calibration session on the NI 9231. The default password is 

NI

.

b) Input the external temperature in degrees Celsius.
c) Call the NI 9231 get C Series adjustment points function to obtain an array of

recommended calibration voltages for the NI 9231.

d) Set the calibrator to a reference value determined by the array of recommended

calibration voltages.

e) Set the calibrator to Operate mode (OPR).
f) Call and configure the NI 9231 adjustment function according to 

NI 9231

Adjustment Configuration

 table.

NI 9231 Calibration Procedure 

  | 

© National Instruments

  | 

5

Summary of Contents for NI 9231

Page 1: ...NI 9231...

Page 2: ...Update 6 Reverification 6 Accuracy under Calibration Conditions 6 Worldwide Support and Services 7 Software Calibrating the NI 9231 requires the installation of NI DAQmx 18 1 or later on the calibrat...

Page 3: ...AQmx NET Help Support for Visual Studio Reference information for NI DAQmx NET methods and NI DAQmx NET properties key concepts and a C enum to NET enum mapping table Test Equipment The following tabl...

Page 4: ...ement circuitry is at a stable operating temperature Initial Setup Complete the following steps to set up the NI 9231 1 Install NI DAQmx 2 Make sure that the NI cDAQ 9178 power source is not connected...

Page 5: ...s the NI 9231 ground connection 3 Set the calibrator voltage to a Test Point value indicated in the NI 9231 Verification Test Limits table 4 Set the calibrator to Operate mode OPR 5 Acquire and averag...

Page 6: ...stment The following performance adjustment procedure describes the sequence of operation required to adjust the NI 9231 Accuracy Adjustment Complete the following procedure to adjust the accuracy per...

Page 7: ...an external calibration setting the C Series calibration temperature and closing the external calibration Reverification Repeat the Verification section to determine the As Left status of the device...

Page 8: ...t ni com services for information about the services NI offers Visit ni com register to register your NI product Product registration facilitates technical support and ensures that you receive importa...

Page 9: ...ments EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy...

Reviews: