| Installation
30
|
Puranity TU+
VWR
Chapter 7
Illustration of drain
4
Locate the cable for level control on
the pure water tank and take it into
the level control connector of the
Puranity TU 12+ panel.
5
Use a AD 8mm/ 0.31 inch tube to
connect the tank overflow fitting
connector with the drain. Route the
AD 8mm/ 0.31 inch tube with a free
gravity fall to the drain. (see
“Illustration of drain” on page 30)
6
Open the feedwater tap and check
that there is no leakage on the
connectors and tubes.
Illustration of drain siphon for inserting the
rinsing water and concentrate outlet tube.
NOTE!
Shorten the rinsing water and concentrate
outlet tube to the required length and route it
to an atmospherically vented drain.
The tubes that run from the Puranity TU+
system and the external tank to the drain
must be routed with a downward slope and
without any kinks or restrictions, as this
would result in backing up of the draining
water.
If a standard drain siphon is in place, the
ends of the tubes must be located at least 20
mm/ 0.75 inch above the drain. Attach the
tubes in such a manner that they remain in
their position.
Step
Action
Figure
Level control
cable
Level control
connection
system panel
Tank
Tank overflow fitting
AD 8mm/ 0.31 inch
tube
Minimum distance
20 mm/0.75 inch
Overflow tube
Rinse water tubing
Drain
Summary of Contents for Puranity TU 12 UV+
Page 6: ...vi Puranity TU VWR...
Page 18: ...Extend of delivery 10 Puranity TU VWR Chapter 3...
Page 20: ...Intended use of the device 12 Puranity TU VWR Chapter 4...
Page 44: ...Flow chart 36 Puranity TU VWR Chapter 8...
Page 46: ...How the Puranity TU system functions 38 Puranity TU VWR Chapter 9...
Page 72: ...Equipment disposal 64 Puranity TU VWR Chapter 14...
Page 76: ...Consumables 68 Puranity TU VWR Chapter 16...
Page 77: ...VWR Puranity TU 69 Terminal assignment...
Page 78: ...Terminal assignment 70 Puranity TU VWR Chapter 17...
Page 80: ...Maintenance record 72 Puranity TU VWR Chapter 18...
Page 84: ...76 Puranity TU VWR Index...