29
Installation |
Chapter 7
Connecting an external tank to the Puranity TU 12+ system
NOTE!
The external pure water tanks are supplied only with the Puranity TU 12+ system.
The external pure water tanks are available in two different sizes:
(171-1170) pure water tank 30L
(171-1171) pure water tank 60L
All connectors for connecting the 30L or 60 L pure water tank to the Puranity TU 12+
system have the same position.
If you ordered the sterile overflow for pure water tanks, connect the sterile overflow as
described in the separate installation guide. If not use a AD 8mm/ 0.31 inch tube to
connect the tank overflow to drain as described in this chapter.
A tank vent filter has to be ordered separately (see “Accessories” on page 67).
Step
Action
Figure
1
Place the external 30L or 60L pure
water tank in the direct vicinity of the
Puranity TU 12+ system.
2
Use the AD 1/4“ tube to connect the
tank back flow connector on the pure
water tank with the tank back flow
connector on the Puranity TU 12+
panel.
3
Plug in the AD 1/4“ tube to connect
the tank flow connector on the pure
water tank to the tank flow connector
on the Puranity TU 12+ panel.
Tank back flow
connection
Puranity TU12+
Tank back flow
connection tank
Tank flow connection
Puranity TU 12+
Tank flow connection
tank
Summary of Contents for Puranity TU 12 UV+
Page 6: ...vi Puranity TU VWR...
Page 18: ...Extend of delivery 10 Puranity TU VWR Chapter 3...
Page 20: ...Intended use of the device 12 Puranity TU VWR Chapter 4...
Page 44: ...Flow chart 36 Puranity TU VWR Chapter 8...
Page 46: ...How the Puranity TU system functions 38 Puranity TU VWR Chapter 9...
Page 72: ...Equipment disposal 64 Puranity TU VWR Chapter 14...
Page 76: ...Consumables 68 Puranity TU VWR Chapter 16...
Page 77: ...VWR Puranity TU 69 Terminal assignment...
Page 78: ...Terminal assignment 70 Puranity TU VWR Chapter 17...
Page 80: ...Maintenance record 72 Puranity TU VWR Chapter 18...
Page 84: ...76 Puranity TU VWR Index...