Instructions for Use
64
Table 21: Sample Method Options
Single Sample
One standard test per sample volume
Control Only
One control test per sample (such as QC check)
C 1 Sample
One baseline control test plus one comparative standard
test per sample
C 2 Samples
One baseline control test plus up to two comparative
standard tests per sample
etc
etc
C 6 Samples
One baseline control test plus up to six comparative
standard tests per sample
This setting determines how the Lu-mini Unit processes the sample tube
readings and displays the result to the user.
7.7.2.1
Single Tube Measurements
For the
Single Sample
and
Control Only
options, the Unit measures a
single test tube and presents the result in the specified numerical format with
optional Banding (see Section 7.7.4).
7.7.2.2
Multiple Tube Measurements
For multiple sample tests (
C 1 Sample
thru
C 6 Samples
),
however, the additional
Sample
results are compared against the baseline
Control
test result and displayed as a bar chart (see Section 7.7.4 for
details).
The first sample test is always on the
Control
sample, and is represented as
normal numeric value (as shown in Section 6.4.4).
When the
Control
tube is removed, the
Tube Selection
screen is displayed.