Instructions for Use
24
6.4
Selecting a Protocol and Performing a Reading
Once the
User ID
has been checked and set, the sample measurement
Protocol
can be selected so that a sample test can be performed.
6.4.1
Selecting a Protocol
The Protocol function allows the user to select the appropriate measurement
protocol for the test about to be performed. To select the measurement
Protocol, press the
Protocol
button from the
Main Menu
.
Protocol
Selection Screen
User ID
Protocol Count
Main
Menu
Screen
→
→
Main
Menu
Worklist Move
Up
Move
Down
Accept
By default, the
Protocol
selection list consists of just the
Standard RLU
Measurement
with no other protocols to choose from, therefore the
Up
and
Down
buttons at the bottom of the screen are greyed-out.
The
Worklist
button is also greyed-out at this point (see Section 7.7.2.3).
There is a small
User ID
icon at the top of the protocol screen to indicate the
currently active User ID.
Additional Protocols and Worklists can only be added and edited
via the Lu-mini Windows Application. See Section 7.7
for details.