Instructions for Use
71
7.7.4.1 Single Sample Result Banding
For
Single Sample
and
Control
sample readings, the quantitative result
value can be qualified by checking it against up to six
Result Bands
. For
instance, the above example translates as:
0 RLU < Result < 500 RLU
LOW
500 RLU < Result < 5000 RLU
NEGATIVE
5000 RLU < Result < 10000 RLU
UNCERTAIN
10000 RLU < Result < 1000000 RLU
POSITIVE
1000000 RLU < Result
OVER RANGE
Ticking the
Limit Result
check box would restrict the test result value to
within the minimum and maximum bands limits of 0 to 1000000 RLU.
Ticking the
Lock Result Status
check box prevents the unit user from
changing the result
Status
tag (see Section 6.4.5).
On the Lu-mini unit Result screen this is displayed as:
Summary of Contents for Lu-Mini
Page 1: ...Instructions for Use...
Page 85: ...Instructions for Use 85 Notes...