Instructions for Use
68
The
Measurement Type
(specified in Table 20) defines how measurement
data is acquired and processed by the Lu-mini unit:
Table 22: Sample Measurement Types
Standard
Simple reading with additional averaging window (T2)
Dual Point (T2-T1)
Difference Reading between readings at T1 and T2
Reaction Rate
As Dual Point but result expressed as
per second
Peak Detect
Peak detection between T1 and T2, with window
average and detection threshold
Kinetic Data
Captures real time readings data for offline analysis.
Refer to Section 7.4.2 for data export options
The
Measurement Type
determines the purpose of the other measurement
parameters:
Measurement
Type
T1
(seconds)
T2
(seconds)
Window
(seconds)
Threshold
(%)
Standard
Averaging
Window
Start
Averaging
Window
End
-
-
Dual Point (T2-T1)
Start
Time
Maximum
Time
-
-
Reaction Rate
Start
Time
Maximum
Time
-
-
Peak Detect
Start
Time
Maximum
Time
Averaging
Window
Percentage
Drop
Kinetic Data
-
Total Data
Points
Sub-data
Rate
-
For
Kinetic Data
capture, the
Window
value defines the number of 1-second
reading to be averaged per outputted data point and
T2
defines to total
number of data points; thus, the total capture time is
T2
*
Window
seconds.
The
Show Graph
option is mainly intended for use when developing a new
Protocol so that the reaction data can be observed in real time. Then, once
the protocol parameters have been finalised, this option can be set to
No
to
hide the graph from the normal end-user.