Instructions for Use
67
7.7.3
Measurement Types
The next group of Protocol definition fields relate to how the sample is
measured by the Lu-mini Unit to produce a test result value.
When the Lu-mini unit measures a test sample, it performs the following
steps, as defined by the selected
Protocol
:
Acquire measurement data
→
Optional: Pre-zero phase
Below
→
Open internal shutter
→
Initial 10-second reading
→
Optional: Additional 1 to 600-second data collection
Below
→
Close internal shutter
Translate reading data according to Assay curve
Process reading data (as per Protocol definition)
Below
Calculate final test result
Hence the measurement sequence is:
→
Luminescent
Signal Data
Acquisition
→
Assay
Standard
Curve
→
Process
Reading
Data
→
Calculate
Formatted
Result
→
Result
Sample
Read
Translate
Process
Calculate
Store
The measurement parameters are:
Measurement Type
Sample measurement type
Show Graph
Show data progress graph for long measurement times
T1
Result analysis start time t1 in seconds
T2
Result analysis end time t2 in seconds
Window
Result analysis window size in seconds
Threshold
Percentage change threshold limit