Instructions for Use
82
11 Technical Specification
Physical Unit Properties
Dimensions (W x D x H)
Weight
160 mm x 235 mm x 140 mm
1.9 kg
Power Supply
Power Adaptor Input Voltage
Power Adaptor Consumption
Power Adaptor Output
Unit Input Voltage and Current
100 to 240V AC ±10%, 50/60 Hz ±5%
25W max
5.0V DC ±5%, 1.0A max
USB-B: 5.0V DC ±5%, 1.0A max
Operating Environment
Temperature Range
Relative Humidity Range
Maximum Operating Altitude
+15 to +35
C
20% to 85% non-condensing
2000 m above sea-level
Storage and Transportation
Temperature Range
Relative Humidity Range
-10 to +50
C
20% to 95% non-condensing
Sample Tube Types
Round Bottomed Tubes
Flat Bottomed Vials
Square Cuvettes
Minimum Sample Volume
10-15mm diameter, 47-75mm tall
15mm diameter, 48mm tall
12.5mm square, 45mm tall
Refer to Table 4
Measurement Range
Standard RLU Protocol
0 to 8x10
6
RLU
Unit Data Storage Capacity
Users
Test Protocols
Test Results
50
100
1000
(or up to 30000 seconds of Kinetic Data)