G4 Multi Channel Weighing Instrument
3-31
Range/Alternatives
Explanation and
<default value>
result of alternatives.
COM2:Mode
Not in use
Modbus Slave
Printer
<Not in use>
Defines the use for serial port Com 2.
Not in use: The serial port is not used.
Modbus Slave: The serial port is used for communication with a
control unit.
Printer: The port is used for a printer.
COM2:Baudrate
300
600
1200
2400
4800
9600
19200
38400
57600
115200
<9600>
Defines the baudrate for the serial communication.
The parameter must be set to the baudrate of
the external equipment.
Note: This parameter is not shown if ‘Not in use’ is selected in
‘COM2:Mode’.
COM2:Data Format
7-none-2
7-even-1
7-even-2
7-odd-1
7-odd-2
8-none-1
8-none-2
8-even-1
8-odd-1
< 8-none-1 >
Defines the bit configuration for the serial communication.
The parameter must be set to the same configuration
as for the external equipment.
Note: This parameter is not shown if ‘Not in use’ is selected in
‘COM2:Mode’.
COM2:Min Reply T
Range: 0 to 1000
Unit: ms
<0>
Adds a delay before the response to a Modbus command is
sent. Used if the instrument sends its response to fast for the
Modbus master.
Note: This parameter is only shown if ‘Modbus Slave’ is selected
in ‘COM2:Mode’.
COM2:Float Form.
Modicon Float
Float
< Modicon Float >
Sets how the Modbus Slave should handled floating point
values.
Modicon Float: Modicon floating point format.
Float: IEEE 32 bit floating point format.
See chapter ‘Communication’ for details on floating point values.
Note: This parameter is only shown if ‘Modbus Slave’ is selected
in ‘COM2:Mode’.
Summary of Contents for G4
Page 2: ......
Page 18: ...Technical Manual 1 14 ...
Page 39: ...G4 Multi Channel Weighing Instrument 3 7 ...
Page 104: ...Technical Manual 5 18 ...
Page 158: ...Technical Manual 6 54 ...
Page 178: ...Technical Manual 7 20 ...
Page 186: ...Technical Manual 8 8 ...
Page 196: ......
Page 197: ...G4 Multi Channel Weighing Instrument Appendix 1 ...
Page 198: ...Technical Manual ...
Page 199: ......