G4 Multi Channel Weighing Instrument
6-11
General registers
The instrument has a number of Modicon 'Holding Registers' (registers 4XXXX...).
The Modbus function 03 'Read Holding Registers' should be used to read these
registers and the Modbus function 05 'Preset Single Register' or 16 'Preset Multiple
Registers' should be used to write to the registers.
See section ‘Data representation’ for a description of the different data formats used.
Hint: To find out which of the float formats that should be used, read
the ‘Instrument type’ register (44000), which equals ‘4001’ for the instrument.
Data type:
Integer
Data type: float
(2 reg./value)
Explanation
R/W
40001 (1 reg)
44000
Instrument type
R
40002 (1 reg)
44002
Standard program major version
R
40003 (1 reg)
44004
Standard program minor version
R
40004 (1 reg)
44006
Special program major version
R
40005 (1 reg)
44008
Special program minor version
R
40006 (3 reg)
44010
Serial number
R
40030 (1 reg)
44030
Command error
R
40031 (1 reg)
44032
Instrument state
R
40032 (1 reg)
44034
Instrument error
R
40033 (1 reg)
44036
Instrument status
R
40034 (1 reg)
44038
Scale 1: Error code
R
40035 (1 reg)
44040
Scale 1: Status
R
40036 (3 reg)
44042
Scale 1: Gross weight
R
40039 (3 reg)
44044
Scale 1: Net weight
R
40042 (3 reg)
44046
Scale 1: Flow rate
R
40045 (3 reg)
44048
Scale 1: Input signal (mV/V)
R
40048 (1 reg)
44050
Scale 2: Error code
R
40049 (1 reg)
44052
Scale 2: Status
R
40050 (3 reg)
44054
Scale 2: Gross weight
R
40053 (3 reg)
44056
Scale 2: Net weight
R
40056 (3 reg)
44058
Scale 2: Flow rate
R
40059 (3 reg)
44060
Scale 2: Input signal (mV/V)
R
Summary of Contents for G4
Page 2: ......
Page 18: ...Technical Manual 1 14 ...
Page 39: ...G4 Multi Channel Weighing Instrument 3 7 ...
Page 104: ...Technical Manual 5 18 ...
Page 158: ...Technical Manual 6 54 ...
Page 178: ...Technical Manual 7 20 ...
Page 186: ...Technical Manual 8 8 ...
Page 196: ......
Page 197: ...G4 Multi Channel Weighing Instrument Appendix 1 ...
Page 198: ...Technical Manual ...
Page 199: ......