G4 Multi Channel Weighing Instrument
3-11
Parameters
On the following pages a survey of all parameters is presented. The parameters are
divided in groups following the menu they belong to. For choice parameters the
available choices are given. For numerical parameters, a value range is given.
At the end of the table, the default value is given in < >.
To the right there is a short parameter explanation and,
in italic,
the results for
the different alternatives.
Range/Alternatives
Explanation and
<default value>
result of alternatives.
Menu ‘General’
Language
English
Svenska
<English>
Defines the language to be used in menus and messages.
Instrument Name
<>
A 16-character string that is used at printing in reports and so
on.
Note: the RM display can’t view all characters that can be set-
up on a graphical screen type of system. The RM can only
enter characters it also can view.
Start Mode
Command
Auto
<Auto>
Defines the start mode after power-on or reset.
Command: A ‘start operation’ command from
control computer or panel key is required for start up.
Auto: Automatic start up.