Technical Manual
8-4
General errors
These errors generally occur due to faulty entries from the front panel, alternatively
invalid data or not allowed commands from the control unit.
Error
code
Explanation
008
Scale not used
The scale that is accessed via communication is not configured (used) in
the instrument.
060 – 063 AOUT ch X termination error
There is a termination error on an analog output channel.
Too low impedance on a voltage output or too high impedance
on a current output may cause this error.
070
I/O module communication error
There is an internal error in the communication between the CPU module
and one or more of the I/O modules.
Restart the instrument and contact the supplier if the error persists.
071 – 074 I/O module reset
There has been an internal reset within an I/O module.
Restart the instrument and contact the supplier if the error persists.
076
Power Fail detected
The instrument have detected that the supply voltage is too low and enters
a power fail state. Weight values etc. are invalid in this state. Power fail
state last a short moment until power is completely lost. If this state
continues more than a moment it might be due to a low input voltage (below
specifications).
100
Instrument in wrong state
The transmitted command is not applicable to the present instrument mode.
101
Too high value!
Value over allowed range. Compare with restrictions for the parameter.
102
Too low value!
Value under allowed range. Compare with restrictions for the parameter.
103
Illegal start address.
Illegal modbus start address, when writing data to the G4 instrument.
104
Illegal number of registers.
Illegal number of modbus registers, when writing data to G4 instrument.
105
Faulty value
Value not accepted when entering a parameter from front panel or illegal
data in modbus registers, when writing data to the G4 instrument.
106
Previous command not ready
The previous command is not finished. Wait until pending command is
finished and send the command again. This error code will only be given via
communication.
110
Unstable weight.
Zero setting always requires a stable weight on the scale. If motion check
is set to ON, also taring and printing of weight values require a stable
weight on the scale. Consequently, if you try to transmit a command for
zero setting, taring, or printing of weight value without awaiting a stable
weight you will receive this error code.
Summary of Contents for G4
Page 2: ......
Page 18: ...Technical Manual 1 14 ...
Page 39: ...G4 Multi Channel Weighing Instrument 3 7 ...
Page 104: ...Technical Manual 5 18 ...
Page 158: ...Technical Manual 6 54 ...
Page 178: ...Technical Manual 7 20 ...
Page 186: ...Technical Manual 8 8 ...
Page 196: ......
Page 197: ...G4 Multi Channel Weighing Instrument Appendix 1 ...
Page 198: ...Technical Manual ...
Page 199: ......