8.3.6.5. Preheating
level
In this menu, it is possible to set the Tubehead pre-heating values. The
value must be set according to the data reported on the "Equipment
parameters table" supplied with the Service Manual.
Once the patient centering (>0<) key is pressed, the following message is
displayed:
I D L E F I L L E V E
L
L
E
V
E
L
=
3 0
NOTE:
DO NOT modify the "LEVEL" value in this step.
Pressing the right key, the next message is displayed:
F I L L E V E
L
@ X
X
m
A
L
E
V
E
L
=
Y
Y
Y
where:
•
"XX" is the Tubehead filament current value, that can be changed
using the left/right keys.
•
"YYY" is the Tubehead pre-heating value, that can be changed with
the increase/decrease keys .
*
Summary of Contents for Rotograph EVO
Page 1: ...Release 14 September 2011 Rev 1 Rotograph EVO 0051 Service Manual ...
Page 3: ...SERVICE MANUAL Revision history Rotograph EVO Rev 1 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 32: ...SERVICE MANUAL Description Rev 1 Rotograph EVO 3 13 Figure 3 1 Block diagram ...
Page 37: ...SERVICE MANUAL Description Rotograph EVO Rev 1 3 18 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 51: ...SERVICE MANUAL Technical data Rotograph EVO Rev 1 4 14 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 59: ...SERVICE MANUAL Pre installation Rotograph EVO Rev 1 5 8 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 169: ...SERVICE MANUAL Troubleshooting Rotograph EVO Rev 1 8 78 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 219: ...SERVICE MANUAL Spare Parts Rotograph EVO Rev 1 10 10 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 222: ......