SERVICE MANUAL
Troubleshooting
Rotograph EVO
(Rev. 1)
8-16
8.2.3.5. E381: Timeout on activating CAN protocol on Generator board /
E382: HF not answering to CAN protocol
These two message have the same meaning, that is a no answer to main
CPU can messages, but their occurring is in different times.
E381 is generated during power on, while E382 is displayed after a
regular power on sequence.
In both cases, the error can be generated by a bad connection of the
CANBus or a fault on the local Generator CPU board or on the Generator
board, so the steps above described for E380 are applied.