SERVICE MANUAL
Troubleshooting
(Rev. 1)
Rotograph EVO
8-17
8.2.4.
Errors with code from E700 to E799
WARNING:
Those errors are related to the X-ray generator, so they can be safety
related.
In case of Error messages E759 and E755, the system must be
immediately powered off, because there is an unexpected emission
(E759) or the emission has not been terminated into the expected
time.
WARNING:
On the Generator board (A10) there are dangerous high voltage,
230 VAC and 400 VDC.
Before accessing the Generator CPU and power board (A9 and A10),
it is mandatory to switch off the mains and wait up to 2 minutes in
order to allow the discharge of the capacitor
.