9
Troubleshooting ALE-NC series operation manual
-
78
-
Error Message/
Error Code
Cause
Coping method
ERR735
Time-out error of importing the sample weight in
the actual value setting method at Percent
weighing mode.
-
-
Improper setting of the weighing pan or
pan base is suspected.
Check for contact with other object.
- Check for any wind or vibration.
ERR736
The setting value is out of the setting range at
numeric value setting method at Percent weighing
mode (lower limit to maximum capacity).
Set the value within the range.
ERR737
-
Sample weight in the air is out of the importing
range at specific gravity mode (over 0g to
maximum capacity).
- Divide the sample so as to its weight in
the air is within the importing range.
-
Sample weight in the water/liquid is out of the
importing range at specific gravity mode (“0 –
maximum capacity” to “maximum capacity”).
- Divide the sample so as to its weight in
the air is within the importing range.
ERR738
Time-out error of importing the sample weight in
the water/liquid at specific gravity mode.
-
-
Improper setting of the weighing pan or
pan base is suspected.
Check for contact with other object.
- Check for any wind or vibration.
ERR746
Invalid date or time was input at <647 DATE
SETTING> or <648 TIME SETTING>.
Set the date and time correctly.
ERR747
Time-out error of importing the sample weight in
the actual value setting method at Comparator
function.
-
-
Improper setting of the weighing pan or
pan base is suspected.
Check for contact with other object.
- Check for any wind or vibration.
ERR748
The setting value is out of the setting range at
numeric value setting method or actual value
setting method at Comparator mode (“0 –
maximum capacity” to “maximum capacity”).
Set the value within the range.
ERR751
The unit weight of the samples is lighter than the
minimum piece weight (MPW) of the balance at
Counting mode.
Choose the samples of which unit
weight is larger than the Minimum Piece
Weight (MPW) of the balance.
ERR752
The unit weight of the samples is 0 g and under at
Counting mode.
-
-
Choose the samples of which unit
weight is larger than the minimum
interval of the balance.
Counting mode cannot operate
subtractive counting.
ERR753
Time-out error of importing the unit weight at
Counting mode.
-
-
Improper setting of the weighing pan or
pan base is suspected.
Check for contact with other object.
ERR754
Deleted the latest data then executed deleting
operation of the second latest data at statistics
mode.
-
-
Only the latest data can be deleted.
Select <<ALL>> to delete all the other
data.
ERR756
Weight of the sample is out of the importing range
at Statistics mode (0g to maximum capacity).
Choose the sample of which weight is
within the importing range.
ERR763
The calculation error of the specific gravity of the
sample at specific gravity mode.
Re-execute the specific gravity function.
ERR764
External weight used for <631 EX CAL> is
different from the selected weight range at
<SELECT WEIGHT>.
Use the external weight of which weight
is within the selected range.