Functions related to the operation ALE-NC series operation manual
-
28
-
3-4
Multiplied by Coefficient mode
Measured weight is multiplied by the preset coefficient, and the result be displayed.
This mode is not available on verified balance.
1
Select the Multiplied by Coefficient mode.
Press [Menu] key, then press [F1]-[F4]
keys to go to <11 MODE>.
Press [F4] key to change the setting
value.
Press [F1]/[F2] key to select.
<<MULT>>
:
Multiplied by Coefficient mode
Press [F4] key to fix.
2
Exit the setting menu.
Press [Menu] key to shift to the
Multiplied by Coefficient mode.
3
Select whether or not employ the previous
recorded coefficient.
Press [F3]/[F4] key to select whether or
not employ the previous data.
When there is no data record, this step is
skipped.
Press [F3]/[F4] key to select.
<<NO>>: Change
<<YES>>: Not Change
When <<OK>> is selected, go to step 6.
4
Set the coefficient.
Numeric value input
(Refer to “2-5-3 Numeric value input”)
Input the coefficient.
Press [F4] key to fix.
5
Weigh the samples.
The weight of the sample is multiplied by the
coefficient and the result is displayed.
3-4-1
Switching the display at Multiplied by Coefficient
1
Press [F1]-[F4] keys to switch the display.