Functions related to the operation ALE-NC series operation manual
-
26
-
3-3
Percentage mode
The weight of a sample to be weighed is indicated in percent relative to the reference weight.
There are two methods to input the reference weight:
- Actual value setting method (<<onW>>):
Place the reference weight on the balance to record the
weight.
- Numeric value setting method (<<NUM>>): Input numeric value of the reference weight by key
operation.
(1) Minimum Reference Weight (MRW)
Models
d (g)
MRW (g)
ALE322NC
0.01
1.00
ALE1501NC
0.1
10.0
ALE8200NC
1
100
(2) The minimum percent to be displayed is automatically set according to the recorded reference
weight.
Readability
(%)
Range of reference weight
1
MRW <=
Reference weight
< MRW x 10
0.1
MRW x 10 <=
Reference weight
< MRW x 100
0.01
MRW x 100 <=
Reference weight
1
Select the percentage mode.
Press [Menu] key, then press [F1]-[F4] keys
to go to <11 MODE>.
Press [F4] key to change the setting value.
Press [F1]/[F2] key to select.
<<PCNT>>
:
Percentage mode
Press [F4] key to fix.
2
Exit the setting menu.
Press [Menu] key to shift to the percentage
mode.
3
Select whether or not employ the previous
recorded reference value.
Press [F3]/[F4] key to select whether or not
employ the previous data.
When there is no data record, this step is
skipped.
Press [F3]/[F4] key to select.
<<NO>>: Change
<<YES>>: Not Change
When <OK> is selected, go to step 6.
4
Select the method of setting the reference
value.
Press [F3]/[F4] key to select.
<<onW>>
:
Actual value
<<NUM>>
:
Numeric value