3
Functions related to the operation ALE-NC series operation manual
-
24
-
3
Place the samples.
Place the set number of samples on the
container.
4
Record the unit weight.
Press [F4] key to fix.
The unit weight is recorded.
5
Simple SCS method
(When enabled).
・・・
When <1C Simple SCS> is valid, Simple
SCS method is activated and the sample
counting indication blinks during this
function.
Add more samples, then the number of
samples and unit weight is automatically
updated when the indication becomes
stable. The number of additional samples
can be up to two times the number of the
samples of the latest update.
For example, when “10 PC” is set, add 20
or less samples.
Repeat this step until the number of the
samples has reached approximately
one-fifth to one-half of the total numbers
that you are intended to count.
Press [F4] key to fix the updated unit
weight.
6
Put objects in place to count them.
Place the objects.
Count result is displayed.
(1) When <on VAR> is selected in step 2, select the specified number of the sample among 10 to 999
by operating [F1]/[F2] keys.
(2) When simple SCS is operating, if the weight of the samples is less than
the “SCS weight”
―
99 times of the minimum readability (d x 99)
―
,
<Add> blinks on the display and unit weight cannot be updated.
In this case, add samples until <Add> indication disappears, or select
the larger number of samples in step 2.
Model
Readability d (g)
SCS weight (g)
ALE322NC
0.01
0.99
ALE1501NC
0.1
9.9
ALE8200NC
1
99
(3) When simple SCS is operating, if the number of the additional samples
is larger than two times of the sample number of the latest update,
<Sub> blinks on the display and unit weight cannot be updated.
In this case, decrease the number of additional samples.