Functions related to the performance ALE-NC series operation manual
-
44
-
4-3
Response speed
The larger numeric value is set in this setting menu, the more stable the balance indication becomes in
unstable conditions.
For verified balance, this function is not available and <22 RESPONSE > is fixed to <<0>>.
1
Select the response speed.
- Press [Menu] key, then press [F1]-[F4]
keys to go to <22 RESPONSE>.
- Press [F4] key to change the setting
value.
- Press [F1]/[F2] key to select.
Refer to Set List.
- Press [F4] key to fix.
Set list
<<0>> : Sensitive mode
<<1>> : Fast
<<2>> : Medium fast
<<3>> : Medium
<<4>> : Medium slow
<<5>> : Slow
2
Exit the setting menu.
- Press [Menu] key to shift to the
measuring mode.
4-4
Automatic zero tracking
The Automatic zero tracking makes it possible to automatically correct the zero-point fluctuation caused
by the temperature fluctuation, etc. when “0” is indicated, through which the “0” indication is maintained.
For verified balance, this function is not available and <23 ZERO TRAC> is fixed to <<OFF>>.
1
Select the automatic zero tracking.
- Press [Menu] key, then press [F1]-[F4]
keys to go to <23 ZERO TRAC>.
- Press [F4] key to change the setting
value.
- Press [F1]/[F2] key to select.
Refer to Set List.
- Press [F4] key to fix.
Set list
<<OFF>> : Invalid
<<0.5>> : 0.5d
<<1>> : 1d
<<2>> : 2d
<<4>> : 4d
2
Exit the setting menu.
- Press [Menu] key to shift to the
measuring mode.