ALE-NC series operation manual 5
Comparator setting
-
47
-
5-2-1
Actual value setting method
1
Set a lower limit value.
Press [F3] key to select.
<<onW>>
:
Actual value
Place a sample to be weighed that is
equivalent to the lower limit value.
Press [F4] key to fix.
The lower limit value is recorded.
2
Set a reference value.
(In the case of the relative value discrimination)
Press [F3] key to select.
<<onW>>
:
Actual value
Place a sample to be weighed that is
equivalent to the reference limit value.
Press [F4] key to fix.
The reference value is recorded.
3
Set an upper limit value.
Press [F3] key to select.
<<onW>>
:
Actual value
Place a sample to be weighed that is
equivalent to the upper limit value.
Press [F4] key to fix.
The upper limit value is recorded.
5-2-2
Numeric value setting method
1
Set a lower limit value.
Numeric value input
(Refer to “2-3-3 Numeric value input”)
Press [F4] key to select.
<<NUM>>
:
Numeric value
Input the lower limit value.
Press [F4] key to fix.
The lower limit value is saved.
2
Set a reference value.
(In the case of the relative value discrimination)
Numeric value input
(Refer to “2-3-3 Numeric value input”)
Press [F4] key to select.
<<NUM>>
:
Numeric value
Input the reference value.
Press [F4] key to fix.
The reference value is saved.
3
Set an upper limit value.
Numeric value input
(Refer to “2-3-3 Numeric value input”)
Press [F4] key to select.
<<NUM>>
:
Numeric value
Input the upper limit value.
Press [F4] key to fix.
The upper limit value is saved.