ALE-NC series operation manual 4
Functions related to the performance
-
43
-
4
Functions related to the performance
Set the parameters relating measurement performance.
For verified balance, these functions are not available.
4-1
Hierarchy of functions related to the performance
- Gray-shaded items (
) are not available on verified balance.
*1 <21 STABLE> is fixed to <<1>>, <22 RESPONSE> is fixed to <<0>>,
and <23 ZERO TRAC> is fixed to <<OFF>> on verified balance.
★
: Initial setting value
Functions relating to
the performance
2 PERFORMANCE
*1
Stability discrimination width
21 STABLE
0.5
±0.5d
*1 ★
1
±1d
2
±2d
4
±4d
*1
Response speed
22 RESPONSE
*1
0
Sensitive mode
1
Fast
*1 ★
2
Medium fast
3
Medium
4
Medium slow
5
Slow
*1
Automatic Zero-Tracking
23 ZERO TRAC
*1
OFF
OFF
*1 ★
0.5
0.5 d/s
1
1 d/s
2
2 d/s
4
4 d/s
4-2
Stability discrimination width
When the larger numeric value is set in this setting menu, the laxer stability judgement is applied, and
the balance indicate “Stable mark” <
> in more unstable conditions.
For verified balance, this function is not available and <21 STABLE> is fixed to <<1>>.
1
Select the stability discrimination width.
- Press [Menu] key, then press [F1]-[F4] keys to
go to <21 STABLE>.
- Press [F4] key to change the setting value.
- Press [F1]/[F2] key to select.
<<0.5>> : 0.5d
<<1>> : 1.0d
<<2>> : 2.0d
<<4>> : 4.0d
- Press [4] key to fix.
2
Exit the setting menu.
- Press [Menu] key to shift to the measuring
mode.