CAA06M User's Guide
23
Quick calibration
The Quick calibration that is available only in the DTF measurement mode is useful when you want to
measure the cable length only, without having to perform a full calibration. It's only valid only for a DTF
measurement in the single layout. If you change the display mode to others, calibration status goes back
to the previous one.
Procedure
1. Connect an extension cable, if necessary, to the
Reflection / RF Out
port of the instrument.
2. Tap
Cal
icon in the right side bar. The on-screen instruction appears containing the
Quick Cal
button.
3. Connect the
OPEN
standard of the calibration kit to the instrument's the
Reflection / RF Out
port or to
the extension cable if used.
4. Tap the
Quick Cal
button. The calibration progress bar appears.
5. Check the screen to see if the calibration status is now changed to
Cal ON (Q)
.
Performing 2-port Calibration
For 2 Port Transmission measurement mode that needs the 2-port calibration, you must perform the
mechanical or electronic O-S-L calibration and then the Thru calibration after setting the frequency and
the output power.
Summary of Contents for CAA06M
Page 2: ...1 CAA06M Cable and Antenna Analyzer User s Guide ...
Page 3: ......
Page 9: ......
Page 17: ...Introduction to CAA06M 8 ...
Page 21: ......
Page 27: ......
Page 42: ...CAA06M User s Guide 33 ...
Page 43: ......
Page 52: ...CAA06M User s Guide 43 DTF VSWR Smith chart measurement screen vertical layout ...
Page 53: ......
Page 59: ...1 Port Cable Loss Testing 50 ...
Page 65: ......
Page 71: ...Smith Chart Testing 62 ...
Page 77: ...2 Port Transmission Testing 68 ...
Page 95: ...Appendix 86 ...