Performing Calibration
20
Procedure
1. Connect an extension cable, if necessary, to the
Reflection / RF Out
port of the instrument.
2. Tap
Cal
icon in the right side bar. The on-screen instruction appears to guide you through the
mechanical O-S-L calibration.
3. Connect the
OPEN
standard of the calibration kit to the instrument's
Reflection /
RF Out
port or to the
extension cable if used.
4. Tap the
O-S-L
button. The calibration progress bar appears.
5. Disconnect the calibration kit and then connect the
SHORT
standard of the calibration kit to the
instrument's the
Reflection / RF Out
port or to the extension cable if used.
6. Tap the
Continue
button. The calibration progress bar appears.
7. Disconnect the calibration kit and then connect the
LOAD
(50 Ω) standard of the calibration kit to the
instrument's the
Reflection / RF Out
port or to the extension cable if used.
8. Tap the
Continue
button. The calibration progress bar appears.
9. Check the screen to see if the calibration status is now changed to
Cal ON
.