Audio/Demod Signal Rx Meter Remote Commands
5 - 17
5.8
AF LEVEL MEASUREMENTS
5.8.1
AF Level - Averages
:CONFigure:AF:ANALyzer:LEVel:AVERage
:CONFigure:AF:ANALyzer:LEVel:AVERage?
5.8.2
AF Level - Average Measurement Reset
:AF:ANALyzer:LEVel:CLEAR:AVG
5.8.3
AF Level - Lower Limit Enable
:LIMits:AF:LEVel:LOWer:ENABle
:LIMits:AF:LEVel:LOWer:ENABle?
Description:
Set command defines the number of readings taken to calculate Average AF
Level measurement.
Query command returns parameter setting.
Range:
1 to 250
Set/Query Format:
NR1
Default Value:
1
Example:
:CONFigure:AF:ANALyzer:LEVel:AVERage 75
Sets number of readings being taken to calculate Average AF Level
measurement to 75.
Query Response:
:CONFigure:AF:ANALyzer:LEVel:AVERage?
75
Description:
Command clears and resets Average AF Level measurement.
Parameter/Query:
none
Description:
Set command Enables/Disables Lower Limit for AF Level measurement.
Query command returns parameter setting.
Parameter:
OFF | ON | 0 | 1
Set/Query Format:
Boolean
Default Value:
OFF
Example:
:LIMits:AF:LEVel:LOWer:ENABle ON
Enables Lower Limit for AF Level measurement.
Query Response:
:LIMits:AF:LEVel:LOWer:ENABle?
1
Summary of Contents for 3900 Series
Page 1: ...3900 Series Digital Radio Test Set dPMR Remote Programming Manual ...
Page 4: ...Preface ii THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 6: ...Contents iv THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 44: ...Generator Transmit Channel Remote Commands 2 34 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 69: ...Analyzer Receiver Remote Commands 3 25 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 147: ...Audio Demod Signal Rx Meter Remote Commands 5 54 THIS PAGE INTENTIONALLY LEFT BLANK ...