Audio/Demod Signal Rx Meter Remote Commands
5 - 16
5.7.10
AF Hum & Noise - Upper Limit Value
:LIMits:AF:HN:UPPer:VALue
:LIMits:AF:HN:UPPer:VALue?
Description:
Set command defines the Upper Limit Value for AF Hum and Noise
measurement.
Query command returns parameter setting.
Range:
-100.0 to +100.0 dBr
Units:
dBr
Default Value:
10.0 dBr
Set/Query Format:
NRf | NR2
Example:
:LIMits:AF:HN:UPPer:VALue 25dBr
Sets Lower Limit Value for AF Hum & Noise measurement to 25.0 dBr.
Query Response:
:LIMits:AF:HN:UPPer:VALue?
25
AF Analyzer Noise measurement type must be defined as Signal to Noise Ratio
(:AF:ANALyzer:NTYPe SN) and SNR measurement defined as Hum & Noise
(:CONFigure:AF:ANALyzer:SNR:MODE 0) to obtain valid Hum & Noise
measurement.
N O T E
Summary of Contents for 3900 Series
Page 1: ...3900 Series Digital Radio Test Set dPMR Remote Programming Manual ...
Page 4: ...Preface ii THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 6: ...Contents iv THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 44: ...Generator Transmit Channel Remote Commands 2 34 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 69: ...Analyzer Receiver Remote Commands 3 25 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 147: ...Audio Demod Signal Rx Meter Remote Commands 5 54 THIS PAGE INTENTIONALLY LEFT BLANK ...