Audio/Demod Signal Rx Meter Remote Commands
5 - 7
5.5
AF DISTORTION MEASUREMENTS
5.5.1
AF Distortion - Averages
:CONFigure:AF:ANALyzer:DISTortion:AVERage
:CONFigure:AF:ANALyzer:DISTortion:AVERage?
5.5.2
AF Distortion - Average Measurement Reset
:AF:ANALyzer:DISTortion:CLEAR:AVG
5.5.3
AF Distortion - Lower Limit Enable
:LIMits:AF:DISTortion:LOWer:ENABle
:LIMits:AF:DISTortion:LOWer:ENABle?
Description:
Set command defines the number of readings taken to calculate Average AF
Distortion measurement.
Query command returns parameter setting.
Range:
1 to 250
Default Value:
1
Set/Query Format:
NR1
Example:
:CONFigure:AF:ANALyzer:DISTortion:AVERage 75
Sets number of readings taken to calculate Average AF Distortion
measurements to 75.
Query Response:
:CONFigure:AF:ANALyzer:DISTortion:AVERage?
75
Description:
Command clears and resets Average AF Distortion measurement.
Parameter/Query:
none
Description:
Set command Enables/Disables Lower Limit for AF Distortion measurement.
Query command returns parameter setting.
Parameter:
OFF | ON | 0 | 1
Default Value:
OFF
Set/Query Format:
Boolean
Example:
:LIMits:AF:DISTortion:LOWer:ENABle ON
Enables Lower Limit for AF Distortion measurement.
Query Response:
:LIMits:AF:DISTortion:LOWer:ENABle?
1
Summary of Contents for 3900 Series
Page 1: ...3900 Series Digital Radio Test Set dPMR Remote Programming Manual ...
Page 4: ...Preface ii THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 6: ...Contents iv THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 44: ...Generator Transmit Channel Remote Commands 2 34 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 69: ...Analyzer Receiver Remote Commands 3 25 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 147: ...Audio Demod Signal Rx Meter Remote Commands 5 54 THIS PAGE INTENTIONALLY LEFT BLANK ...