dPMR Signal Rx Meter Remote Commands
4 - 21
4.9.8
Symbol Clock Error - Upper Limit Enable
:LIMits:SCE:CH1:UPPer:ENABle
:LIMits:SCE:CH1:UPPer:ENABle?
4.9.9
Symbol Clock Error - Upper Limit Value
:LIMits:SCE:CH1:UPPer:VALue
:LIMits:SCE:CH1:UPPer:VALue?
Description:
Set command Enables/Disables Upper Limit for Symbol Clock Error
measurement.
Query command returns parameter setting.
Parameter:
OFF | ON | 0 | 1
Default Value:
OFF
Set/Query Format:
Boolean
Example:
:LIMits:SCE:CH1:UPPer:ENABle ON
Enables Upper Limit for Symbol Clock Error measurement.
Query Response:
:LIMits:SCE:CH1:UPPer:ENABle?
1
Description:
Set command defines Upper Limit Value for Symbol Clock Error measurement.
Query command returns parameter setting.
Range:
0.0 to +1000.0 mHz
Units:
mHz
Default Value:
0.0 mHz
Set/Query Format:
NRf | NR2
Example:
:LIMits:SCE:CH1:UPPer:VALue 500mHz
Sets Upper Limit Value for Symbol Clock Error measurement to 500.0 mHz.
Query Response:
:LIMits:SCE:CH1:UPPer:VALue?
500.00
Summary of Contents for 3900 Series
Page 1: ...3900 Series Digital Radio Test Set dPMR Remote Programming Manual ...
Page 4: ...Preface ii THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 6: ...Contents iv THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 44: ...Generator Transmit Channel Remote Commands 2 34 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 69: ...Analyzer Receiver Remote Commands 3 25 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 147: ...Audio Demod Signal Rx Meter Remote Commands 5 54 THIS PAGE INTENTIONALLY LEFT BLANK ...