Operation, Remote Operation
Measurement Commands
Rev. L0
ATC-5000NG ATC/DME Test Set Operation Manual
1-2-5
February 2020
Subject to Export Control. See Cover Page for details.
Page 7
5.3
MEASUREMENT COMMANDS
This set of commands allows the user to query the ATC-5000NG to perform measurements on the
received signal from the UUT. The ATC-5000NG can perform frequency, power and pulse characteristic
measurements on both 1030 and 1090 MHz signals.
5.3.1
DATA FORMAT
5.3.2
DF DATA FRAME BIT
5.3.3
MODE
Command Syntax:
{:ATC |:ATC5000NG}{:MEA |:MEASURE}{:DFORMAT}SP<numeric>CR
Description:
This command sets the replies numeric format.
Numeric:
0 to 2 (decimal ASCII)
Value
0
1
2
Format
Hexadecimal
Decimal (no fractional part)
Float (fractional part)
Example:
:ATC:MEA:DFORMAT 2\r
Default:
1
Query:
:ATC:MEA:DFORMAT?\r
Return:
0
Command Syntax:
{:ATC |:ATC5000NG}{:MEA |:MEASURE}{:SET |:SETTINGS}{:DFBIT}SP<numeric>CR
Description:
This command sets the measurement routine to sample the selected bit.
Numeric:
1 to 112 (decimal ASCII)
Example:
:ATC:MEA:SET:DFBIT 7\r
Default:
1
Query:
:ATC:MEA:SET:DFBIT?\r
Return:
7
Command Syntax:
{:ATC |:ATC5000NG}{:MEA |:MEASURE}{:SET |:SETTINGS}{:MOD |:MODE}SP<numeric>CR
Description:
This command sets the measurement mode. This command is ignored when the scenario type
selected is XPDR.
Numeric:
0 (decimal ASCII)
Value
0
Measurement Mode
Pulse
Example:
:ATC:MEA:SET:MOD 0\r
Default:
Pulse
Query:
:ATC:MEA:SET:MOD?\r
Return:
0