Operation, Remote Operation
Measurement Commands
Rev. L0
ATC-5000NG ATC/DME Test Set Operation Manual
1-2-5
February 2020
Subject to Export Control. See Cover Page for details.
Page 9
5.3.5
PULSE REQUEST
5.3.5.1
FREQUENCY REQUEST
5.3.5.2
PULSE FALLTIME REQUEST
5.3.5.3
PULSE RISETIME REQUEST
5.3.5.4
PULSE POSITION REQUEST
5.3.5.5
PULSE POWER REQUEST
Command Syntax:
{:ATC |:ATC5000NG}{:MEA |:MEASURE}{:FREQ? |:FREQUENCY?} CR
Description:
This command returns the frequency of the incoming signal.
Example:
:ATC:MEA:FREQ?\r
Return:
Value is in ASCII data format specified in MHz. In case of the Float data format, the Value is
returned with three decimal points.
If no measurement Value is available, the Unit replies with not ready “#”.
Command Syntax:
{:ATC |:ATC5000NG}{:MEA |:MEASURE}{:PUL |:PULSE}{:FALL?} CR
Description:
This command returns the fall time of the selected pulse.
Example:
:ATC:MEA:PUL:FALL?\r
Return:
Value is in ASCII data format specified in ns. In case of the Float data format, the Value is
returned without fractional part.
If no measurement Value is available, the Unit replies with not ready “#”.
Command Syntax:
{:ATC |:ATC5000NG}{:MEA |:MEASURE}{:PUL |:PULSE}{:RISE?} CR
Description:
This command returns the rise time of the selected pulse.
Example:
:ATC:MEA:PUL:RISE?\r
Return:
Value is in ASCII data format specified in ns. In case of the Float data format, the Value is
returned without fractional part.
If no measurement Value is available, the Unit replies with not ready “#”.
Command Syntax:
{:ATC |:ATC5000NG}{:MEA |:MEASURE}{:PUL |:PULSE}{:POS?|:POSITION?} CR
Description:
This command returns the position of the selected pulse.
Example:
:ATC:MEA:PUL:POS?\r
Return:
Value is in ASCII data format specified in ns. In case of the Float data format, the Value is
returned without fractional part.
If no measurement Value is available, the Unit replies with not ready “#”.
Command Syntax:
{:ATC |:ATC5000NG}{:MEA |:MEASURE}{:PUL |:PULSE}{:POWER?} CR
Description:
This command returns the power of the selected pulse.
Example:
:ATC:MEA:PUL:POWER?\r
Return:
Value is in ASCII data format specified in dBm. In case of the Float data format, the Value is
returned with two decimal points.
If no measurement Value is available, the Unit replies with not ready “#”.
Summary of Contents for 138156