Diagnostics and repair
LFXG-H Installation and Operation Guide
79
Analog alarm
If the current loop output (analog output) is stable at 2 mA or 22 mA, the
analog alarm is set.
The analog alarm is set when the counts from the detector falls below a set
threshold, indicating that the detector is not outputting enough counts to
make a meaningful measurement. This is known as 0 counts.
If the analog alarm is on, verify:
Source holder shutter is in the On or Open position to create the
required radiation field.
Extreme build-up on walls or other material shielding the detector from
the radiation field.
Damage or disconnection of electrical connections from the sensor
assembly to the CPU board.
New hardware
found? – No/Yes
New hardware found – The
CPU board detects a
configuration mismatch. The
CPU board or sensor
assembly may have been
replaced, or one of the
EEPROM configurations is
erroneous.
Contact VEGA Field
Service. See 26.
Process out of
range? – No/Yes
Process out of measurement
range – The current process
value is not within the limits
set by the Max level and Min
level in the gauge span
settings.
Call VEGA Field
Service.
X-Ray Alarm – No/
Yes
Note that there are high levels
of x-ray in your area that may
be affecting process
measurement.
Contact VEGA for
further information.
Table 5.3
Diagnostic alarm conditions (continued)
Diagnostic check
and Normal/Error
conditions
HART message
Diagnostic description
Action
Summary of Contents for FiberFlex LFXG-H
Page 2: ......
Page 4: ...ii LFXG H Installation and Operation Guide NOTES...
Page 6: ...iv LFXG H Installation and Operation Guide...
Page 14: ...Preface xii LFXG H Installation and Operation Guide...
Page 34: ...Installation 32 LFXG H Installation and Operation Guide FiberFlex mounting...
Page 91: ...Diagnostics and repair LFXG H Installation and Operation Guide 89 Source Functions...
Page 100: ...98 LSGH Installation and Operation Guide NOTES...
Page 101: ......