Advanced functions
68
LFXG-H Installation and Operation Guide
Test
Current Loop Test (milliamp output)
This mode manually forces the current output to a specified value. This is
useful for verifying the current loop calibration. To calibrate the current loop,
see Chapter 3: “Calibration”.
Procedure 4.5: To perform a current loop test
ailable:
1.
Select
Diagnostics | Test | Current Loop Test
.
2.
Click
Enter
.
3.
Remove the gauge from control.
4.
Enter the current loop test value.
5.
Click
OK
.
The transmitter functions in this mode until it times out (1 hour), or
you click
Exit
and
OK
.