Advanced functions
LFXG-H Installation and Operation Guide
61
Percent Count Range
The compensated measurement counts that express
as a percent of the counts at the high and low-
endpoints of the calibration (determined with the two
point calibration.) This quantity shows where the
current measurement is in relation to the total count
range.
% count range = 100 x (C
L
- C
S
) / (C
L
- C
H
)
where
C
S
= sum counts
C
L
,C
H
= counts at Cal Low level and Cal High level
C
L
-C
H
= counts range
Percent Process Span
The measurement value as a percent of the
measurement span. Enter the maximum and minimum
level values in the Setup tab.
A graph of percent count range vs. percent process
span indicates the non-linearity of the radiation
transmission’s measurement. If using a table linearizer,
the values in the table are percent count range and
percent process span.
Unfiltered PV
The level in inches without the time constant or
rectangular window filter.
Uncompensated PV
The level of process before any process compensation.
Final PV
The process value in engineering units after applying
the filter.This value relates to the current loop output.
Aux Counts
The frequency-input counts from optional auxiliary
input.
Filtered Aux Counts
The filtered auxiliary counts. Enter the filter dampening
value for the auxiliary input’s filter time-constant.
Table 4.1
Process Chain tab — display values (continued)
Value
Description
Summary of Contents for FiberFlex LFXG-H
Page 2: ......
Page 4: ...ii LFXG H Installation and Operation Guide NOTES...
Page 6: ...iv LFXG H Installation and Operation Guide...
Page 14: ...Preface xii LFXG H Installation and Operation Guide...
Page 34: ...Installation 32 LFXG H Installation and Operation Guide FiberFlex mounting...
Page 91: ...Diagnostics and repair LFXG H Installation and Operation Guide 89 Source Functions...
Page 100: ...98 LSGH Installation and Operation Guide NOTES...
Page 101: ......