Calibration
42
LFXG-H Installation and Operation Guide
To correlate the 4 ... 20 mA to the process value, set the span of the current
loop output.
A direct measurement of the current is preferable: hook the meter up in
series with the instrument and the DCS. However, if you know the
resistance of the DCS, use a voltage measurement to calculate the current.
Measuring the current loop output
Before a current loop calibration:
Connect an ammeter or the DCS to:
•
Terminal connections 13 (mA +) and 14 (mA -)
•
Test points H1 and H2
•
Anywhere along the current loop
Make sure there is a 250 – 800
load on the current loop. If no load or
an insufficient load exists on the loop, it may require temporary
placement of a resistor across terminals 13 and 14. Hook the meter or
DCS in series with the load resistor.
Note:
The current loop and process spans are independent and set
separately. The current loop span sets the level indications for
the 4 mA and the 20 mA outputs. The process span sets the
endpoints of the calibration curve.
Current meter
Voltmeter
Terminal block pins
13 and 14
Detector
housing
DCS
R
t
R
t
Terminal block pins
13 and 14
Detector
housing
Summary of Contents for FiberFlex LFXG-H
Page 2: ......
Page 4: ...ii LFXG H Installation and Operation Guide NOTES...
Page 6: ...iv LFXG H Installation and Operation Guide...
Page 14: ...Preface xii LFXG H Installation and Operation Guide...
Page 34: ...Installation 32 LFXG H Installation and Operation Guide FiberFlex mounting...
Page 91: ...Diagnostics and repair LFXG H Installation and Operation Guide 89 Source Functions...
Page 100: ...98 LSGH Installation and Operation Guide NOTES...
Page 101: ......